4610-- WATER PURFICATION SYSTEM HCC1 LAB

Agency:
State: Federal
Level of Government: Federal
Category:
  • 46 - Water Purification and Sewage Treatment Equipment
Opps ID: NBD00159412417419253
Posted Date: Feb 29, 2024
Due Date: Mar 7, 2024
Source: Members Only
Follow
4610-- WATER PURFICATION SYSTEM HCC1 LAB
Active
Contract Opportunity
Notice ID
36C24624Q0414
Related Notice
36C24624Q0414
Department/Ind. Agency
VETERANS AFFAIRS, DEPARTMENT OF
Sub-tier
VETERANS AFFAIRS, DEPARTMENT OF
Office
246-NETWORK CONTRACTING OFFICE 6 (36C246)
Looking for contract opportunity help?

APEX Accelerators are an official government contracting resource for small businesses. Find your local APEX Accelerator (opens in new window) for free government expertise related to contract opportunities.

APEX Accelerators are funded in part through a cooperative agreement with the Department of Defense.

The APEX Accelerators program was formerly known as the Procurement Technical Assistance Program (opens in new window) (PTAP).

General Information
  • Contract Opportunity Type: Solicitation (Original)
  • All Dates/Times are: (UTC-05:00) EASTERN STANDARD TIME, NEW YORK, USA
  • Original Published Date: Feb 29, 2024 11:02 am EST
  • Original Date Offers Due: Mar 07, 2024 10:00 am EST
  • Inactive Policy: Manual
  • Original Inactive Date: Mar 12, 2024
  • Initiative:
    • None
Classification
  • Original Set Aside:
  • Product Service Code: 4610 - WATER PURIFICATION EQUIPMENT
  • NAICS Code:
    • 333310 - Commercial and Service Industry Machinery Manufacturing
  • Place of Performance:
Description
VA Chesapeake Community Based Outpatient Clinic (CBOC)
Chesapeake HCC1

Salient Characteristics

Water Purification System that is capable of producing not less than 200 liters of water per hour, fitted with deionization capabilities

MODEL MEDICA-R 200
Make-up rate @15°C 200 l/hr
Delivery flow rate maximum115V, 60Hz 21 l/min @ 45psi (3bar) Inorganics (resistivity @25°C) >10 M -cm (1)
Organics (TOC) - typical Particles 0.2 m filter Silica Notes: 1 with Ion-exchange cylinder installed (Nuclear or Hypex grade resin), 2 subject to correct operating and maintenance procedures
Feedwater Requirement:
Source: Tap water as detailed below Conductivity: Contaminant:
Measure: Range: Pretreatment:
Calcium: Ca ppm as CaCO3 >250 Softener or use very low RO recovery3
Free chlorine: Cl2ppm 0.1 0.5 20 inch carbon block
>0.5 Cylinder of carbon sized correctly to obtain Chloramine: Cl2 ppm 0.1 0.5 20 inch high efficiency carbon (5)
>0.5 Cylinder of carbon sized correctly to obtain Silica: SiO2 ppm >30 20 inch cartridge depth filter or use very low RO recovery (4)
Fouling index: FI 10 to 20 20 inch cartridge depth filter
>20 Back washable media filter with a minimum flow rate of 20 l/min
Iron/manganese: Fe/Mn ppm >0.05 20 inch cartridge depth filter (5)
>0.1 Back-washable Fe filter (6)
Organics: TOC ppm C 2 to 3 20 inch carbon block (4)
- >3 Cylinder of carbon sized correctly for TOC demand (4) Temperature: 4 - 40°C (Recommended 15 - 25°C)
-
Flowrate: (requirement at 15°C) 20 l/min
Drain requirements: (gravity fall with air gap) 45 l/min
Feedwater pressure: 60psi (4bar) maximum, 30psi (2bar) minimum
Notes: 3) Check LSI, increase frequency of acid cleaning, 4) Increase frequency of alkaline cleaning,
5) Please use ELGA Part Numbers LA683 (Filter Housing) and LC172 (Carbon Filter), 6) Increase frequency of acid cleaning Wall mount canister filter will need (Wall backing required)
Must include: Installation, Commissioning, Training
Electrical Requirements Mains Input: 115V, 60Hz
System control voltage (not including pumps and UV) 24V dc Power consumption (peak demand): 2000VA
Electrical protection rating 20 amps Noise level during recirculation ***Note: System requires approximately 4' x 6' of floor space for tanks, filtration Unit, and filters.
-The floor drain must be a minimum of 3" and meet all applicable building codes and system requirements
-Actual Unit measurement 32" Wide x35"Deep x72"High

VA Chesapeake Community Based Outpatient Clinic (CBOC)
Chesapeake HCC1

Page 10 of 10

-

-
-
- Qty: 1
- Finishes/Options
- Image








PATHOLOGY AND LABORATORY
LAB-C1012(1)
Department, Room, Quantity
Attachments/Links